Fig. 6 | Journal of Cardiovascular Magnetic Resonance

Fig. 6

From: 2D cine DENSE with low encoding frequencies accurately quantifies cardiac mechanics with improved image characteristics

Fig. 6

Phase noise had a larger effect on displacement errors with lower ke. a In a stationary water phantom, phase noise, as quantified by RMSE in radians, was inversely related to SNR. No differences in RMSE were seen between the different ke. b When RMSE in radians was converted to millimeters by dividing by ke, there were substantial differences between the different ke. Lower ke had increased displacement errors