Skip to main content

Table 5 Reproducibility of Black Blood T2* at 1.5 T and 3 T

From: Comparison of 3 T and 1.5 T for T2* magnetic resonance of tissue iron

Test Geometric Mean (95 % CI) of Squared difference P-value
  T2* BB 1.5 T T2* BB 3 T  
Intra-Observer 0.04 (0.02, 0.09) 0.07 (0.03, 0.15) 0.26
Inter-Observer 0.04 (0.01, 0.12) 0.07 (0.02, 0.21) 0.18
Inter study 0.26 (0.08, 0.80) 0.28 (0.09, 0.82) 0.94