Table 3 Short-term reproducibility experiments in a 3 T final phantom (E-model)*

From: A medical device-grade T1 and ECV phantom for global T1 mapping quality assurance—the T1 Mapping and ECV Standardization in cardiovascular magnetic resonance (T1MES) program

Tube

Parameter

Sequence

CoV (%)

Mean diff. ± s.d.

A

T1

pre_MOLLI_5(3)3_256_T1

0.16

255 ± 0.4

 

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.18

255 ± 0.5

T2

T2_4pt_TRUFI_192i_T2

0.66

194 ± 1.3

T2_4pt_GRE_192i_T2

0.61

134 ± 0.8

J

T1

pre_MOLLI_5(3)3_256_T1

0.14

1860 ± 2.6

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.17

1672 ± 2.8

T2

T2_4pt_TRUFI_192i_T2

4.06

227 ± 9.2

T2_4pt_GRE_192i_T2

1.37

203 ± 2.8

C

T1

pre_MOLLI_5(3)3_256_T1

0.08

460 ± 0.4

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.08

461 ± 0.4

T2

T2_4pt_TRUFI_192i_T2

0.52

195 ± 1.0

T2_4pt_GRE_192i_T2

0.76

160 ± 1.2

K

T1

pre_MOLLI_5(3)3_256_T1

0.13

953 ± 1.2

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.10

917 ± 0.9

T2

T2_4pt_TRUFI_192i_T2

0.98

60 ± 0.6

T2_4pt_GRE_192i_T2

0.67

49 ± 0.3

L

T1

pre_MOLLI_5(3)3_256_T1

0.08

1372 ± 1.1

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.16

1252 ± 2.0

T2

T2_4pt_TRUFI_192i_T2

0.91

56 ± 0.5

T2_4pt_GRE_192i_T2

0.89

49 ± 0.4

M

T1

pre_MOLLI_5(3)3_256_T1

0.15

1178 ± 1.8

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.12

1104 ± 1.3

T2

T2_4pt_TRUFI_192i_T2

0.91

58 ± 0.5

T2_4pt_GRE_192i_T2

0.66

49 ± 0.3

G

T1

pre_MOLLI_5(3)3_256_T1

0.19

285 ± 0.6

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.20

285 ± 0.6

T2

T2_4pt_TRUFI_192i_T2

0.29

86 ± 0.2

T2_4pt_GRE_192i_T2

1.02

49 ± 0.5

H

T1

pre_MOLLI_5(3)3_256_T1

0.11

527 ± 0.6

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.09

527 ± 0.5

T2

T2_4pt_TRUFI_192i_T2

0.35

66 ± 0.2

T2_4pt_GRE_192i_T2

0.72

46 ± 0.3

I

T1

pre_MOLLI_5(3)3_256_T1

0.06

406 ± 0.3

post_MOLLI_4(1)3(1)2_256_MOCO_T1

0.05

409 ± 0.2

T2

T2_4pt_TRUFI_192i_T2

0.21

72 ± 0.2

T2_4pt_GRE_192i_T2

0.19

47 ± 0.1

  1. *All tests performed at 20.3 ± 0.48 °C on Siemens, Skyra 3 T at RBHT, November 2015 with RR interval 900 ms and using two T1 mapping sequences (pre-MOLLI 5(3)3 [256] and post-MOLLI 4(1)3(1)2 [256] with MOCO, WIPs# 780B) and two T2 mapping sequences (TRUFI T2 map and GRE T2 map)
  2. CoV coefficients of variation, diff. difference, GRE gradient echo, MOCO motion correction, MOLLI modified Look-Locker inversion recovery, RR inter-beat interval, s.d. standard deviation, TRUFI true fast imaging with steady-state free precession