Skip to main content

Table 2 Imaging parameters of BTI, MPRAGE and CE-MRV in the patient study

From: Black-blood thrombus imaging (BTI): a contrast-free cardiovascular magnetic resonance approach for the diagnosis of non-acute deep vein thrombosis

Parameters

BTI

MPRAGE

CE-MRV

Repetition time, ms

650

1500

2.8

Echo time, ms

9.8

1.7

1.6

Flip angle

variable

18

18

Fat suppression

fat saturation

water excitation

fat saturation

GRAPPA in the phase-encoding direction

2

off

3

Turbo factor

40

288

NA

Slice Partial Fourier

6/8

off

6/8

Phase Partial Fourier

off

off

6/8

Bandwidth, Hz/Pixel

710

685

685

True resolution, mm3

1.1 × 1.1 × (1.1–1.3)

1.3 × 1.3 × 1.3

1.1 × 1.1 × 1.2

Reconstructed resolution, mm3

0.55 × 0.55 × (0.55/0.65)

0.65 × 0.65 × 0.65

0.55 × 0.55 × 0.60

Field of view, mm2

352 × 352

340 × 340

400 × 400

Partition

208–256

288

192

Scan time per sequence

3 min 23 s–4 min 49 s

6 min 23 s

27 s

Inversion time, ms

NA

400

NA

  1. Note: GRAPPA Generalized Autocalibrating Partially Parallel Acquisitions, NA Not Applicable