Skip to main content

Table 1 Protocol parameters for AIF imaging sequence at 1.5 T

From: Myocardial perfusion cardiovascular magnetic resonance: optimized dual sequence and reconstruction for quantification

 

FLASH

TE

0.76 & 1.76 ms

TR

2.45 ms

FA

5°

Matrix

64×34

FOV (typical)

360x270x10 mm3

PE order

Linear

Parallel imaging

TPAT2

TI

23.8 ms

SR prep

6-pulse (26 ms incl. spoilers)

Imaging duration

42 ms

Total duration

68.2 ms