Skip to main content

Table 2 Protocol parameters for myocardial perfusion CMR sequence at 1.5 T

From: Myocardial perfusion cardiovascular magnetic resonance: optimized dual sequence and reconstruction for quantification

 

FLASH

SSFP

PD frames

3

PD FA

5° (FLASH)

TE

1.0 ms

1.04 ms

TR

2.1 ms

2.5 ms

Bandwidth

1085 Hz/pixel

FA

14°

50°

Matrix

192×111 (1.9×2.4 mm2)

Partial Fourier

3/4

Asymmetric echo

weak

FOV (typical)

360x270x8 mm3

PE order

Linear

Parallel imaging

TPAT3

TS/TD

100/62 ms

95/40 ms

SR prep

6-pulse (26 ms including spoilers)

Fat saturation

optional

Imaging duration

59 ms

70 ms

Total duration

143 ms/slice

142 ms/slice

3 slices + AIF

497 ms (>120 bpm)

495 ms (>120 bpm)