Table 1 CMR sequence parameters

From: Cardiovascular magnetic resonance imaging in the prospective, population-based, Hamburg City Health cohort study: objectives and design

Modality Pulse sequence Voxel size (mm3) FoV (mm2) TR (ms) TE (ms) FA (°) PAT
Morphology T2 HASTE 1.3 × 1.3 × 8 400 1100 90 110 2
Cine CMR bSSFP 1.6 × 1.6 × 8 340 48 1.5 80 3
Native T1 Mapping MOLLI 5b(3b)3b 1.4 × 1.4 × 8 360 281 1.1 35 2
T2 Mapping T2 prepared FLASH 1.9 × 1.9 × 8 360 207 1.3 12 2
Perfusion CMR Turbo FLASH 1.9 × 1.9 × 8 360 158 1 10 2
Angiography VIBE 1.4 × 1.4 × 5 380 3.3 1.3 9 4
Flow measurements VENC 1.8 × 1.8 × 6 340 46 2.5 20 2
AD and PWV VENC 1.2 × 1.2 × 8 300 17 2.3 20 2
LGE 2D-PSIR 1.6 × 1.6 × 8 400 700 1.2 55 2
PC T1 Mapping MOLLI 4b(1b)3b(1b)2b 1.4 × 1.4 × 8 360 361 1.1 35 2
  1. Abbreviations: AD Aortic Distensibility, FA flip angle, FLASH fast low angle shot, FoV reconstructed field of view, HASTE Half fourier-acquired single shot turbo spin Echo, LAX long axis view, LGE late gadolinium enhancement, MOLLI modified Look-Locker inversion recovery, PAT parallel acquisition technique, PC post-contrast, PSIR phase sensitive inversion recovery, PWV pulse wave velocity, SAX short axis view, bSSFP balanced steady state free precession, TE echo time, TR repetition time, VENC velocity encoded, VIBE volumetric interpolated breath-hold examination