Table 3 Primary observer; ventricular metrics measured using real-time and standard BH imaging

From: Real-time assessment of right and left ventricular volumes and function in children using high spatiotemporal resolution spiral bSSFP with compressed sensing

  Real-timea Standard Breath Holda Bias Limits of agreement
LV EDV (mL) 110 ± 38 112 ± 38b 1.7 ± 4.6 −7.3 to 10.7
LV ESV (mL) 40 ± 16 39 ± 16b − 1.0 ± 3.5 −7.9 to 5.9
LV SV (mL) 70 ± 25 73 ± 24b 2.6 ± 4.8 −6.7 to 12.0
LV EF (%) 64 ± 6 66 ± 7b 1.5 ± 3.0 −4.3 to 7.3
LV mass 88 ± 36 85 ± 34b − 3.1 ± 7.0 − 16.9 to 10.6
RV EDV (mL) 121 ± 37 121 ± 38 0.5 ± 5.2 −9.6 to 10.6
RV ESV (mL) 51 ± 23 49 ± 24 − 1.2 ± 5.0 −11.0 to 8.5
RV SV (mL) 70 ± 20 72 ± 20b 1.8 ± 4.9 −7.8 to 11.4
RV EF (%) 59 ± 9 61 ± 10b 1.4 ± 3.7 −6.0 to 8.7
  1. EDV, end-diastolic volume; EF, ejection fraction; ESV, end-systolic volume, LV, left ventricular; RV, right ventricular; SV, stroke volume
  2. aDisplayed as mean ± standard deviation
  3. b Indicates significant differences with Standard BH technique (p < 0.05)
  4. Bias is the mean of the paired difference (Standard BH – Real-time) presented with standard deviation (SD) of the difference
  5. Limits of agreements are bias ± 1.96xSD