Table 3 Intraobserver variability for software A, B and C

From: Quantification in cardiovascular magnetic resonance: agreement of software from three different vendors on assessment of left ventricular function, 2D flow and parametric mapping

 

Software A

Software B

Software C

Left ventricular assessment

 Ejection Fraction [%]

−0.2 ± 2.4

−0.5 ± 2.7

0.5 ± 1.6

 Mass [g]

2.9 ± 6.8

− 3.4 ± 5.0

− 1.3 ± 3.4

 End-diastolic volume [ml]

1.9 ± 5.0

0.6 ± 5.5

− 0.1 ± 3.4

 End-systolic volume [ml]

0.6 ± 3.7

0.6 ± 2.9

0.3 ± 2.3

2D Flow measurement

 Vmax [cm/s]

0.0 ± 0.0

0.0 ± 0.1

0.0 ± 0.0

 Stroke Volume [ml]

−1.0 ± 2.1

−1.2 ± 2.3

0.0 ± 1.6

Parametric mapping

 T1-time [ms]

−9.4 ± 9.0

−4.0 ± 12.5

−0.6 ± 12.2

 T2-time [ms]

− 0.4 ± 0.9

−0.1 ± 1.6

−0.1 ± 1.1

  1. Intraobserver variability expressed as mean bias ± standard deviation. The greatest standard deviation among the three software is indicated in bold
  2. EF ejection fraction, EDV end-diastolic volume, ESV end-systolic volume, Vmax peak velocity, SV stroke volume