From: A cardiovascular magnetic resonance (CMR) safe metal braided catheter design for interventional CMR at 1.5 T: freedom from radiofrequency induced heating and preserved mechanical performance
Sequence Type
Scan Time
Segmented Braiding (Δ°C)
Non-segmented Braiding (Δ°C)
Real-time bSSFP
60 s
0.78
52.6
20 min
0.92
54.8
Real-time GRE
0.67
13.4
0.82
14.1