Table 1 RF induced temperature rise at the distal tip

From: A cardiovascular magnetic resonance (CMR) safe metal braided catheter design for interventional CMR at 1.5 T: freedom from radiofrequency induced heating and preserved mechanical performance

Sequence Type Scan Time Segmented Braiding (Δ°C) Non-segmented Braiding (Δ°C)
Real-time bSSFP 60 s 0.78 52.6
Real-time bSSFP 20 min 0.92 54.8
Real-time GRE 60 s 0.67 13.4
Real-time GRE 20 min 0.82 14.1