Skip to main content

Table 1 Reference Acquisition Parameters for Contrast Enhanced Cardiovascular Magnetic Resonance (CE-MRA), 3D whole-heart Balanced State Free Precession (bSSFP) and modified Dixon (mDixon) sequences

From: mDixon ECG-gated 3-dimensional cardiovascular magnetic resonance angiography in patients with congenital cardiovascular disease

Parameter

CE-CMRA

bSSFP

mDixon

Average Acquisition Time Goal

Dynamic scan time goal ~ 12 s

3:11 min

3:08 min

TE (ms)

0.87

2

1st image: 1.8

2nd image: 3.4

TR (ms)

2.2

4.5

5.2

Flip Angle (degrees)

30°

90°

15°

FOV (mm)

260 × 260

260 × 260

260 × 260

Acquired Voxel Size

1.4 × 1.4 × 2.8

1.4 × 1.4 × 1

1.4 × 1.4 × 1

Recon Voxel (mm3)

0.9 × 0.9 × 0.4

0.9 × 0.9 × 1

0.9 × 0.9 × 1

Sense

LR: 2.5; AP: 1.0

LR: 1.5; AP: 1.0

LR: 2.0; AP: 1.5

Trigger Delay

NA

Mid-Diastole

Mid-Diastole

Shot Duration (ms)

NA

130

130

Navigator Window (mm)

Cartesian k-space acquisition window

7

7

Slice number

90

135

135

Bandwidth (Hz/pixel units)

0.165

0.35

0.3

  1. Abbreviations: mDixon modified-Dixon, bSSFP balanced Steady State Free Precession, CE-CMRA Contrast-Enhanced Cardiovascular Magnetic Resonance Angiography; TE echo time; TR repetition time; NA not applicable; LR Left Right; AP Anterior Posterior