Table 1 Temperature (°C) increase during defibrillation measured in 10-s averages (N = 6)

From: High-energy external defibrillation and transcutaneous pacing during MRI: feasibility and safety

Temperature

Mean

Std

P (Wilcoxon matched pairs test)

Outside bore

36.35

1.57

 

Inside bore

36.88

2.51

0.010

Scanning

37.03

2.45

0.036