Fig. 1 | Journal of Cardiovascular Magnetic Resonance

Fig. 1

From: Dynamic quantitative nonenhanced magnetic resonance angiography of the abdominal aorta and lower extremities using cine fast interrupted steady-state in combination with arterial spin labeling: a feasibility study

Fig. 1

a Bland-Altman diagram showing the plot of the difference between the flow velocity measured by cine fast interrupted steady state in combination with arterial spin labeling (cFASL) and 2D phase contrast (PC) at all stations in both cohorts against the mean velocity of the pair (n = 38). Red lines show the 95% limits of agreement, and the green line shows the mean value of the differences. b Bland-Altman diagram showing the plot of the difference between the flow velocity measured by cFASL and 2DPC at all stations in the lower extremity volunteers against the mean velocity of the pair (n = 27). Red lines show the 95% limits of agreement, and the green line shows the mean value of the differences. c Bland-Altman diagram showing the plot of the difference between the flow velocity measured by cFASL and 2DPC in the suprarenal aorta against the mean velocity of the pair (n = 11). Red lines show the 95% limits of agreement, and the green line shows the mean value of the differences