Skip to main content

Table 1 T1 relaxation times (ms) for an experimental phantom and simulated heart rates

From: Model-based myocardial T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH cardiovascular magnetic resonance

Tube IR spin-echo T2 [35] HR 0 HR 40 HR 60 HR 80 HR 100
1 315 ± 6 101 ± 2 308 ± 5 309 ± 5 308 ± 5 307 ± 5 310 ± 6
2 497 ± 8 46 ± 2 452 ± 5 453 ± 5 449 ± 5 448 ± 5 452 ± 7
3 661 ± 8 81 ± 3 622 ± 6 622 ± 5 618 ± 5 615 ± 5 617 ± 7
4 822 ± 12 132 ± 5 793 ± 7 792 ± 9 786 ± 8 787 ± 9 793 ± 10
5 1191 ± 13 138 ± 4 1149 ± 15 1140 ± 15 1148 ± 13 1153 ± 15 1159 ± 17
6 1508 ± 15 166 ± 5 1474 ± 14 1462 ± 16 1479 ± 17 1474 ± 17 1480 ± 21
  1. HR Heart rate, IR Inversion recovery