Table 3 The relationship between heart rate and T1 mapping measurements

From: The dynamics of extracellular gadolinium-based contrast agent excretion into pleural and pericardial effusions quantified by T1 mapping cardiovascular magnetic resonance

RR (ms)

HR (bpm)

T1 (ms)

Curve fit uncertainty in SD (ms)

T1* (ms)

Relaxation (%)

600

100

3363

1000

1000

98%

700

86

2948

882

1530

96%

800

75

2853

464

1949

94%

900

67

2808

256

2198

95%

1000

60

2818

158

2389

95%

1100

55

2827

106

2498

96%

1200

50

2859

73

2604

97%

1300

46

2867

54

2612

97%

1400

43

2878

41

2645

98%

1500

40

2910

33

2674

98%

  1. The T1 values of a phantom with a reference T1 value of 3011 ms imaged using MOLLI acquisitions with increasing heart rates (HR). The curve fit uncertainty was obtained from the SD maps provided by the MOLLI reconstruction, T1* is the apparent T1 relaxation time constant and relaxation the percentage of relaxation after 8 RR-intervals, immediately before the second inversion of the MOLLI sequence