Skip to main content

Table 4 Inter-observer reproducibility and effect of 20- vs. 50-time reconstructed frames temporal resolution

From: Comparison of different methods for the estimation of aortic pulse wave velocity from 4D flow cardiovascular magnetic resonance

 

Inter-observer reproducibility

Effect of temporal resolution

Mean bias (m/s)

LA (m/s)

p

ICC

Mean bias (m/s)

LA (m/s)

p

ICC

S1

TTc

−0.50

[−2.6;1.6]

0.59

0.95

−1.62

[−11;8.1]

0.49

0.47

TTw

0.46

[−5.2;6.1]

1.00

0.86

0.29

[−1;1.5]

0.50

0.91

TTf

−0.46

[−3.4;2.4]

0.43

0.91

0.35

[−0.5;1.2]

0.50

0.93

S2

TTc

0.06

[−0.6;0.7]

0.84

0.99

−0.96

[−5.5;3.6]

0.54

0.75

TTw

0.09

[−0.7;0.8]

0.80

0.99

0.15

[−0.5;0.8]

0.76

0.97

TTf

−0.53

[−4.2;3.2]

0.38

0.61

0.15

[−0.6;0.9]

0.92

0.97

S3

0.18

[−0.9;1.2]

0.84

0.97

0.33

[−1.4;2]

0.68

0.82

  1. Measures of repeatability of 4D flow MRI aoPWV estimates are provided: Bland-Altman mean biases and limits of agreement (LA) as well as intra class correlation coefficients (ICC). p values correspond to Wilcoxon rank-sum test of comparisons between aoPWV values obtained by the two observers and for the two reconstructions. S1, S2 and S3: Strategies 1, 2 and 3. TTc, TTw and TTf: transit times estimated using cross-correlation, wavelets and Fourier transforms, respectively