Skip to main content

Table 4 Interstudy bias, interstudy variability, and correlation coefficient

From: A comparison of cine CMR imaging at 0.55 T and 1.5 T

 

Inter study reproducibility (bias ± 1.96xSD) between field strengths

coefficient of variation

Correlation coefficient

LVEDV

All

0.4 ± 18.6 mL (− 18.4 mL to 18.8 mL)

3.3%

0.99

Breath-held

0.0 ± 20.6 mL (−20.6 mL to 20.6 mL)

4.5%

0.98

Free-breathing

0.9 ± 15.9 mL (−15.0 mL to 16.9 mL)

2.3%

0.99

LVESV

All

1.3 ± 14.8 mL (−13.5 mL to 16.2 mL)

5.3%

0.98

Breath-held

1.3 ± 18.2 mL (−16.9 mL to 19.5 mL)

6.4%

0.98

Free-breathing

1.4 ± 9.0 mL (−7.7 mL to 10.4 mL)

3.7%

0.99

LVED Mass

All

0.4 ± 11.2 g (−10.8 g to 11.5 g)

2.9%

0.99

Breath-held

0.1 ± 12.9 g (−12.8 g to 12.9 g)

3.2%

0.99

Free-breathing

0.7 ± 8.6 g (−7.9 g to 9.3 g)

2.5%

0.99

LVES Mass

All

1.3 ± 13.2 g (−11.8 g to 14.6 g)

3.0%

0.99

Breath-held

 2.2 ± 14.9 g (−12.7 g to 17.1 g)

3.6%

0.99

Free-breathing

0.2 ± 10.4 g (−10.2 g to 10.5 g)

2.3%

0.99

LVSV

All

−1.0 ± 17.1 mL (−18.0 mL to 16.1 mL)

5.1%

0.95

Breath-held

−1.2 ± 19.4 mL (− 10.6 mL to 18.2 mL)

6.2%

0.89

Free-breathing

-0.7 ± 13.8 mL (−14.4 mL to 13.1 mL)

3.6%

0.98

LVEF

All

−0.8 ± 7.2% (−8.0 to 6.4%)

5.8%

0.91

Breath-held

−0.9 ± 8.9%(−9.8 to 8%)

6.3%

0.91

Free-breathing

−0.6 ± 4.15% (−4.8 to 3.5%)

5.1%

0.88

RVEDV

All

−1.6 ± 18.5 mL (−20 mL to 16.9 mL)

2.9%

0.98

Breath-held

-1.6 ± 16.0 mL (−17.6 mL to 14.4 mL)

2.6%

0.95

Free-breathing

−1.5 ± 21.7 mL (−23.1 mL to 20.2 mL)

3.3%

0.98

RVESV

All

−1.2 ± 11.7 mL (− 12.9 mL to 10.5 mL)

5.4%

0.97

Breath-held

−0.5 ± 11.7 mL (− 12.2 mL to 11.2 mL)

5.1%

0.95

Free-breathing

−2.2 ± 11.7 mL (−13.9 mL to 9.5 mL)

5.8%

0.98

RVSV

All

−0.2 ± 19.7 mL (−19.9 mL to 19.5 mL)

5.7%

0.92

Breath-held

0.1 ± 17.7 mL (−18.3 mL to 17.1 mL)

5.4%

0.86

Free-breathing

0.3 ± 22.4 mL (−22.1 mL to 22.7 mL)

6.0%

0.94

RVEF

All

−0.1 ± 7.7% (−8.0 to 7.8%)

4.0%

0.82

Breath-held

−0.6 ± 7.4% (−7.5 to 7.3%)

3.9%

0.97

Free-breathing

0.0 ± 8.6% (−8.6 to 8.6%)

4.2%

0.69

  1. Interstudy bias, interstudy variability, and correlation coefficient between 0.55 T and 1.5 T for quantitative ventricular volume and systolic function measurements. Coefficient of variation was calculated from the standard deviation between 0.55 T and 1.5 T measurements, divided by the mean of the two measurements