Skip to main content

Table 1 Sequence parameters; either automatic or volume shim was performed. For (b)SSFP sequences a SENSE factor of 2 was used

From: Clinical experience regarding safety and diagnostic value of cardiovascular magnetic resonance in patients with a subcutaneous implanted cardioverter/defibrillator (S-ICD) at 1.5 T

 

Flip angle in °

Spatial resolution in mm3

Receiver bandwidth

in MHz

TR / TE in msec

Inversion bandwidth in Hz

Acquisition time in sec

bSSFP

60

1.8 × 2.0 × 8

954

3.3 / 1.6

 

5.5

SSFP

15

2.0 × 2.6 × 8

1033

5.3 / 3.1

 

8.5

LGE

25

1.6 × 2.3 × 10

1055

6.1 / 3.0

1000

10.0

wb-LGE

25

1.6 × 2.3 × 10

1055

6.2 / 3.0

4000

10.0

  1. bSSFP balanced steady state free precession; LGE late gadolinium enhancement; wb-LGE wide band