Skip to main content

Table 3 Test-retest reliability, interrater and intrarater agreement of global rest native T1, stress T1 and ΔT1

From: The reliability and feasibility of non-contrast adenosine stress cardiovascular magnetic resonance T1 mapping in patients on haemodialysis

Parameter

Study 1

Study 2

CoV

ICC (95% CI)

Bias ± SD of bias

BA Limits of Agreement

Test-retest reliability (n = 10)

 Rest T1

1273 ± 48

1264 ± 41

1.2%

0.79 (0.38; 0.94)

9.4 ± 29.3

−48.0; 66.8

 Stress T1*

1322 ± 47

1305 ± 53

1.5%

0.69 (0.16; 0.92)

16.4 ± 38.3

−58.6; 91.5

 ΔT1*

3.6 ± 1.8

3.0 ± 2.0

27.4%

0.55 (− 0.10; 0.88)

0.6 ± 1.8

−3.0; 4.1

Interrater agreement (n = 15)

 Rest T1

1272 ± 40

1274 ± 37

0.3%

0.98 (0.95; 0.99)

−1.5 ± 7.0

− 15.2; 12.3

 Stress T1

1313 ± 44

1316 ± 46

0.3%

0.99 (0.96; 1.00)

− 2.4 ± 6.9

− 16.0; 11.2

 ΔT1

3.2 ± 1.9

3.3 ± 1.7

9.6%

0.94 (0.83; 0.98)

−0.1 ± 0.6

−1.3; 1.2

Intrarater agreement (n = 15)

 Rest T1

1272 ± 40

1274 ± 38

0.2%

0.99 (0.97; 1.00)

−1.4 ± 5.4

−12.0; 9.2

 Stress T1

1313 ± 44

1316 ± 46

0.3%

0.98 (0.95; 0.99)

− 2.5 ± 8.5

− 19.2; 14.2

 ΔT1

3.2 ± 1.9

3.3 ± 1.8

10.3%

0.94 (0.83; 0.98)

−0.1 ± 0.7

−1.4; 1.3

  1. Data presented as mean ± SD. * n = 9. CoV Coefficient of variability, ICC Intraclass correlation coefficient, BA Bland-Altman