Parameter | FS (T) | Vendor | Technique | n | Meanp | SDp | LL–ULm |
---|
Native T1 time (ms) | 1.5 | Siemens | MOLLI | 417a | 972 | 43 | 885–1059 |
1.5 | Siemens | ShMOLLI | 971b | 960 | 29 | 903–1017 |
1.5 | Philips | MOLLI | 215c | 989 | 42 | 905–1073 |
3 | Siemens | MOLLI | 301d | 1196 | 47 | 1103–1290 |
3 | Siemens | ShMOLLI | 197e | 1130 | 55 | 1021–1240 |
3 | Philips | MOLLI | 201f | 1097 | 66 | 964–1230 |
ECV (%) | 1.5 | Siemens | MOLLI | 199 g | 26 | 3 | 20–32 |
1.5 | Siemens | ShMOLLI | 295 h | 27 | 3 | 21–33 |
1.5 | Philips | MOLLI | 56i | 23 | 3 | 17–29 |
3 | Siemens | MOLLI | 129j | 26 | 3 | 20–32 |
3 | Siemens | ShMOLLI | 57 k | 25 | 2 | 20–29 |
3 | Philips | MOLLI | 100 l | 26 | 5 | 16–36 |
- ECV extracellular volume fraction, FS field strength, T Tesla, n number of study subjects included in the weighted mean values, meanp pooled weighted mean, SDp pooled standard deviation, LL lower limit, UL upper limit, MOLLI modified look locker inversion-recovery, ShMOLLI short MOLLI, Siemens Siemens Medical Solutions, Erlangen, Germany, Philips Philips Healthcare, Best, The Netherlands
- aPooled weighted values from references [118, 121, 129, 132, 136, 141, 142]
- bPooled weighted values from references [116, 122,123,124, 128, 130, 131, 135, 142, 147, 148]
- cPooled weighted values from references [127, 139, 145, 146]
- dPooled weighted values from references [125, 133, 136, 144, 149]
- ePooled weighted values from references [137, 138]
- fPooled weighted values from references [126, 134, 140, 145]
- gPooled weighted values from references [121, 129, 142]
- hPooled weighted values from references [124, 128, 130, 131, 142]
- iValues from reference [127]
- jPooled weighted values from references [143, 144, 149]
- kValues from reference [137]
- lPooled weighted values from references [139, 140]
- mCalculated as meanp ± 2*SDp