Skip to main content

Table 2 CMR sequence parameters

From: Comparison between conventional and compressed sensing cine cardiovascular magnetic resonance for feature tracking global circumferential strain assessment

 

Conventional cine

CS cine

Sequence type

2D cine TrueFISP

2D cine TrueFISP

ECG mode

Retrospective gating

Prospective triggering

Repetition time (ms)

3.16

3.16

Echo time (ms)

1.4

1.4

FOV (mm2)

350 × 350

350 × 350

Image matrix

208 × 166

208 × 166

Spatial resolution reconstructed (mm2)

1.7 × 1.7

1.7 × 1.7

Temporal resolution (ms) (repetition time × segments)

41.1

41.1

Slice thickness (mm)

6

6

Slice numbers

8

8

Flip angle (degrees)

50

50

Bandwidth (Hz/pixel)

1145

1145

Number of segments

13

13

Cardiac phases

25

19–31

Breath-holds (n)

4

1

Acceleration factor

3

12.8

Iterative reconstruction (n)

-

80

  1. CS compressed sensing, ECG electrocardiogram, FOV field of view