Skip to main content

Table 4 Averaged vessel diameter measurements of both reviewers on non-contrast-enhanced REACT-CMRA, contrast-enhanced steady-state-, and first-pass-CMRA

From: Free-breathing non-contrast flow-independent cardiovascular magnetic resonance angiography using cardiac gated, magnetization-prepared 3D Dixon method: assessment of thoracic vasculature in congenital heart disease

 

REACT–CMRA

Steady–state–CMRA

First–pass–CMRA

P value

Ascending aorta (mm)

28.2 ± 6.1

27.7 ± 6.0

27.9 ± 6.3

0.96

Left pulmonary artery (mm)

18.3 ± 6.0

17.7 ± 5.5

18.8 ± 5.5

0.84

Left superior pulmonary vein (mm)

13.2 ± 5.2

13.1 ± 4.8

12.7 ± 4.4

0.80

  1. Vessel diameters are given as mean ± standard deviation. P values were obtained using one-way ANOVA followed by Tukey multiple comparison tests