Table 1 Typical imaging parameters for various PSIR LGE MOCO protocols

From: Bright-blood and dark-blood phase sensitive inversion recovery late gadolinium enhancement and T1 and T2 maps in a single free-breathing scan: an all-in-one approach

 

Bright Blood (BB)

Dark Blood (DB)

mSASHA (joint T1 and T2 map)

Preparation

Inversion Preparation

Inversion Preparation & T2 preparation

Saturation Preparation & T2 preparations

(see Fig. 1)

Readout (single shot)

SSFP

FAIR = 50°

FAPD = 8°

SSFP

FAIR = 50°

FAPD = 8°

SSFP

Variable FA (max 100°)

Typical FOV / resolution

360 × 270 mm2

1.4 × 1.9 × 8 mm3

Matrix size

256 × 144 (parallel imaging factor 2 & PF factor 7/8)

Number of acquired images

8 (16 beats)

16 (32 beats)

30 (45 heart beats)

T2 prep TE

n/a

Variable (10—40 ms)

55 ms

TE/TR

1.2/2.8 ms

1.2/2.8 ms

1.3/2.9 ms