Table 3 Comparison of device parameters before, after CMR and at 1-year follow-up

From: Feasibility of adenosine stress cardiovascular magnetic resonance perfusion imaging in patients with MR-conditional transvenous permanent pacemakers and defibrillators

 

Before CMR

After CMR

At 1-y follow-up

P value

P-wave amplitude (mV)

4.6 ± 4.4

4.6 ± 4.4

3.3 ± 1.0

0.36

R-Wave amplitude (mV)

11.8 ± 4.8

11.8 ± 4.8

11.3 ± 5.1

0.77

Atrial lead impedance (Ohm)

483.3 ± 96.3

453.9 ± 99.05

484.1 ± 108.2

0.78

Ventricular lead impedance (Ohm)

476.2 ± 106.2

497.3 ± 68.8

464 ± 76.2

0.75

Atrial PCT (V@0.4 ms)

0.72 ± 0.29

0.75 ± 0.24

1.08 ± 1.24

0.22

Ventricular PCT (V@0.4 ms)

0.80 ± 0.21

0.82 ± 0.20

0.84 ± 0.30

0.68