Skip to main content

Table 3 Minimum and Optimal Hardware and Pulse Sequences for CMR

From: 30-minute CMR for common clinical indications: a Society for Cardiovascular Magnetic Resonance white paper

Gradient system

Chest receive coil

Pulse sequences

Minimum

Optimal

Minimum

Optimal

Minimum

Optimal

Amplitudes > 33 mT/m

Slew rates > 120 mT/m/ms

Amplitudes ≥ 40 mT/m

Slew rates ≥ 200 mT/m/ms

8

16 or higher

Localizer

Cine Imaging

Perfusion (for stress)

LGE

Minimum plus -

T1 and T2 mapping

Velocity-encoded cine