CMR technique | CMR sequence | Scan prescription | Scan time |
---|---|---|---|
Localizer | Single shot FSE, GRE, or bight blood bSSFP | 3 orthogonal slices (axial, coronal, sagittal) Transverse stack covering the heart Standard cardiac planes: 2-chamber, 4-chamber, 3-chamber, short axis stack | 5 min |
Perfusion | Saturation recovery bSSFP or GRE | Short axis planes (base, mid, apex); Long-axis plane of choice Simultaneous intravenous injection of contrast (rate = 3–7 mL/s; 30 mL saline flush) | 5 min |
Cine Imaging | k-space segmented cine bSSFP or k-space segmented cine GRE e.g., to reduce susceptibility artifact | Standard cardiac planes: 2-chamber, 4-chamber, 3-chamber, LVOT, and shot axis stack (2D slices covering the heart from base to apex) Breath hold at end-expiration | 8–10 min |
LGE | 2D inversion recovery (IR) bSSFP or GRE; optimal to include phase-sensitive inversion recovery reconstructions | Select inversion time that nulls normal myocardium, optimally with a mid-short axis plane TI scout acquisition Standard cardiac planes (same as cine imaging) | 5–10 min |