Fig. 4 | Journal of Cardiovascular Magnetic Resonance

Fig. 4

From: Validation of 4D flow cardiovascular magnetic resonanceĀ in TIPS stent grafts using a 3D-printed flow phantom

Fig. 4

Comparison of low and high velocity encoded 4D flow CMR of the TIPS stent graft. 4D flow CMR performed at the maximum flow rate of 2.8Ā l/min resulted in flow acceleration within the TIPS stent graft above 1Ā m/s. A Velocity encoding at 1Ā m/s resulted in aliasing within the TIPS stent graft with consecutive erroneous calculation of pathlines as indicated by pathlines perpendicular to the fluid flow (arrow). BĀ Velocity encoding at 2Ā m/s showed no aliasing within the TIPS stent graft and resulted in correct calculation of velocity-coded pathlines. Note that aliasing with erroneous calculation of pathlines results also in erroneous quantification of flow rates and flow velocities. SMVā€‰ā€‰superior mesenteric vein, SVā€‰ā€‰splenic vein, PVā€‰ā€‰portal vein