Fig. 7 | Journal of Cardiovascular Magnetic Resonance

Fig. 7

From: Validation of 4D flow cardiovascular magnetic resonanceĀ in TIPS stent grafts using a 3D-printed flow phantom

Fig. 7

Intra- and interobserver agreement of 4D flow CMR-derived flow rate and velocity measurements. A, B Blandā€“Altman plots of intraobserver agreement (0.9ā€‰Ā±ā€‰5.9%) and interobserver agreement (0.2ā€‰Ā±ā€‰6.4%) for measurements of flow rates within the covered part of the TIPS stent graft (TIPSp, TIPSm, and TIPSv) at velocity encoding of 2Ā m/s. C, D Blandā€“Altman plots of intraobserver agreement (0.4ā€‰Ā±ā€‰1.6%) and interobserver agreement (0.5ā€‰Ā±ā€‰1.5%) for measurements of flow velocities within the covered part of the TIPS stent graft at velocity encoding of 2Ā m/s. Middle solid line indicates mean bias of flow rate and flow velocity measurements. Dashed lines indicate limits of agreement