From: Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance

Fig. 6

Short-term reproducibility of T2 at 1.5 T (left) and 3 T (right) acquired using T2 mapping bSSFP repeated 3 times in each of the final prototypes #Ci (at a temperature of 22°) (n = 1) and #Cii (at 21°) (n = 1) manufactured months apart, from independent stock solutions. All these scans were performed on the same day with independent placement of phantom and shims. Coefficients of variation (CoV) of T2 are shown per tube and were all < 1% in the absence of temperature correction. CoV for T1 using 3 MOLLI repeats are not shown here but were also < 1% for both prototypes (1.5 T range: 0.13–0.94%; 3 T range: 0.03–0.38%)