Table 3 Long-term reproducibility results for one phantom at baseline and its 12 months repeat scan

From: Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance

 

1.5 T n = 1

3 T n = 1

Biological scope (Tube ID)

Baseline

12-month repeat

(absolute difference in ms, % diff)

Baseline

12-month repeat

(absolute difference in ms, % dif)

Short T2 native myocardium at both 1.5 T and 3 T (A)

43

43 (0, 0%)

43

43 (0, 0%)

Medium T2 native myocardium at 1.5 T (B)

49

49 (0, 0%)

48

49 (1, 2.1%)

Medium T2 native myocardium at 3 T (C)

47

47 (0, 0%)

45

45 (0, 0%)

Long T2 native myocardium at 1.5 T (D)

65

66 (1, 1.5%)

65

65 (0, 0%)

Long T2 native myocardium at 3 T (E)

65

65 (0, 0%)

65

65 (0, 0%)

Long-normal T2 native myocardium at 1.5 T (F)

57

57 (0, 0%)

57

57 (0, 0%)

Very long T2 native myocardium at 3 T (G)

74

73 (1, 1.4%)

72

71 (1, 1.4%)

Mildly long T1 & T2 native myocardium at 1.5 & 3 T (H)

57

57 (0, 0%)

56

56 (0, 0%)

Medium T1 & T2 post-GBCA* myocardium at 1.5 & 3 T (I)

56

57 (1, 1.8%)

56

57 (1, 1.8%)

  1. All the T2 presented in this table were acquired using T2 mapping bSSFP in final prototype #Ci at baseline and prototype #Cii at the 12-month repeat
  2. Other abbreviations as in Table 2