Reproducibility of slice-interleaved T1 (STONE) mapping sequence

Background

Slice interleaved T1 mapping sequence (STONE) allows quantification of native T1 of the entire ventricle in a single free-breathing scan. Using this sequence, data acquisition for different slices are interleaved, allowing longer recovery time after the inversion pulse, which could result in improved accuracy and precision. However, measurement reproducibility of the STONE sequence has not been previously studied. In this study, we sought to assess native T1 measurement reproducibility a) within session, b) between sessions and c) between days.

Methods

Eleven healthy subjects (33 ± 16 years, 6 male) underwent non-contrast CMR imaging on 2 different days. Figure 1 shows the study design. Each subject was imaged twice with identical imaging protocol. After image prescription, the subjects were imaged using STONE SSFP with the following imaging parameters: In-plane resolution = 2.1 × 2.1 mm2, slice thickness = 8 mm, slice gap = 4 mm, Field of View = 320 × 320 mm2, TR/TE/α = 2.8 msec. / 1.38 msec. /70°, SENSE-rate = 2, linear ordering, 10 linear ramp-up pulses and acquisition window = 218.8 msec, bandwidth = 1879.7 Hz/pixel. STONE GRE sequence was acquired with following parameters: In-plane resolution = 2 × 2 mm2, slice thickness = 8 mm, slice gap = 4 mm, Field of View = 300 × 300 mm2, TR/TE/α = 3.9 msec. / 1.94 msec. /90°, SENSE-rate = 2.5, linear ordering, 10 linear ramp-up pulses and acquisition window = 166.6 msec., bandwidth= 1315.8 Hz/pixel. Imaging was repeated twice for each sequence. Subsequently, subjects were removed from the scanner and repositioned, followed by the same scan protocol. The same imaging protocol was repeated on a second day of scan. All imaging was performed in a 1.5T CMR scanner (Philips Achieva) using a 32-channel cardiac receiver coil array. T1 maps were created by voxel-wise fitting using a 2-parameter fit model after motion correction. The epicardial and endocardial contours in the left ventricle were manually drawn in 5 short axis-slices to calculate global and slice-based myocardial T1 values. Coefficient of variation analysis for each slice was generated to assess the variability within each session, between the sessions, between different days and for each subject.

Figure 1
figure 1

Study protocol to assess within session, between session and between day (test/re-test) variability for native myocardial T1.

Results

Figure 2 shows mean T1 values for different imaging sessions, averaged over all subjects for STONE-SSFP and STONE-GRE. The CVs for all slices and subjects (figure 2) showed low variability for STONE-SSFP (2.4 ± 1.3%) and for STONE-GRE (1.65 ± 0.95%). The CVs for all slices and days showed a mean of 2.1 ± 1.45% for STONE SSFP and a mean of 1.5 ± 1.1% for Stone GRE. The CVs for all slices and sessions showed a mean of 1.7 ± 1.95% for STONE SSFP and a mean of 1.2 ± 1.3% for STONE GRE.

Figure 2
figure 2

Mean native T1 per repetition, session and day for STONE and coefficient of variation for different slices and subjects.

Conclusions

Native myocardial T1 measurements by STONE-GRE and STONE-SSFP are very reproducible. These data suggest that STONE-SSFP and STONE-GRE should be considered for longitudinal studies to assess potential temporal changes in native T1 values for disease monitoring and/or during therapy.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bellm, S., Basha, T., Ngo, L. et al. Reproducibility of slice-interleaved T1 (STONE) mapping sequence. J Cardiovasc Magn Reson 18 (Suppl 1), P51 (2016). https://doi.org/10.1186/1532-429X-18-S1-P51

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-18-S1-P51

Keywords