Skip to main content

Table 3 Comparison of 2D reference against 3D 7× k-t SENSE and 3D 10× k-t SENSE images (n = 6).

From: Measurement of left ventricular dimensions with contrast-enhanced three-dimensional cine imaging facilitated by k-t SENSE

 

LV EDV (ml)

LV ESV (ml)

LV mass (g)

LVEF (%)

2D mean ± SD

145.8 ± 25.9

63.3 ± 19.1

95.4 ± 18.1

57.2 ± 5.9

3D 7× k-t mean ± SD

138.3 ± 26.2

62.3 ± 20.5

90.9 ± 18.6

60.9 ± 10.7

Bias (95% C.I.)

7.4 (-14.6; 29.4)

1.1 (-12.8; 14.9)

4.4 (-14.4; 23.3)

-3.7 (-26.7; 19.4)

Lin's coefficient (95% C.I.)

0.87 (0.64–1.00)

0.96 (0.88–1.00)

0.84 (0.56–1.00)

0.88 (0.67–1.00)

3D 10× k-t mean ± SD

134.3 ± 25.8

63.9 ± 22.0

92.7 ± 17.2

53.7 ± 8.0

Bias (95% C.I.)

11.5 (-16.1; 39.0)

-0.6 (-20.1; 18.9)

2.7 (-13.5; 18.9)

3.6 (-1.7; 8.8)

Lin's coefficient (95% C.I.)

0.77 (0.41–1.00)

0.89 (0.69–1.00)

0.88 (0.67–1.00)

0.81 (0.58–1.00)