Skip to main content

Table 1 Statistical results

From: Accurate quantification of heart valve regurgitation in all four heart valves simultaneously using 3D velocity-encoded MRI with retrospective valve tracking

  

MV-AV

MV-TV

TV-PV

Volunteers (n = 14)

Pearson correlation

0.96

0.97

0.96

 

p-value t-test

0.01

0.31

0.01

 

Mean difference

6 ± 7 ml

-2 ± 5 ml

6 ± 6 ml

 

Confidence interval

-7 – 19 ml

-12 – 9 ml

-6 – 17 ml

Patients (n = 23)

Pearson correlation

0.95

0.98

0.88

 

p-value t-test

0.07

0.75

0.30

 

Mean difference

3 ± 6 ml

0.3 ± 4 ml

2 ± 10 ml

 

Confidence interval

-10 – 16 ml

-8 – 9 ml

-17 – 22 ml