Contrast enhanced coronary MRA at 3.0 T using a 32 channel coil to detect coronary artery stenosis - how does it measure up against 64-slice coronary CTA and X-ray angiography?
Journal of Cardiovascular Magnetic Resonance volume 12, Article number: P52 (2010)
Introduction
3.0 T contrast enhanced whole-heart coronary MRA has shown high sensitivity for the detection of stenoses in patients suspected of coronary artery disease (CAD). Several studies reported that coronary MRA might have similar accuracy in detecting CAD as 64-Slice CTA. However, no direct comparison between both techniques has been performed.
Purpose
The purpose of this study was to perform a head-to-head comparison of 3.0 T whole-heart coronary MRA using 32 channel coils in patients with suspected CAD compared to 64-slice CTA and x-ray coronary angiography (CAG).
Materials and methods
32 patients with suspected CAD who were scheduled for CAG underwent both coronary MRA and coronary CTA. 3.0 T contrast enhanced whole-heart coronary MRA was acquired by using an ECG-triggered, navigator-gated, inversion-recovery prepared, segmented gradient-echo sequence and 32 channel coils was used with GRAPPA factor of 3. Coronary CTA was performed using a 64-Slice CT scanner. For image analysis, standard 15-segment AHA classification system was used. The diagnostic accuracy of MRA and CTA in detecting significant stenoses (≥ 50% of vessel lumen) was compared on the per-segment basis using CAG as reference standard.
Results
MRA and CTA examinations were successfully completed in all 32 patients. The averaged imaging time of MRA was 6.1 ± 1.2 min. In a total of 419 coronary segments with a reference luminal diameter ≥ 1.5 mm on QCA, forty-nine (12%) on MRA were evaluated as non-assessable and twenty-eight on CTA were non-diagnostic. On the assessable segments basis, 3.0 T whole-heart CMRA correctly identified 45 significant stenoses, and CTA correctly identified 52 significant stenoses. The sensitivity, specificity, PPV and NPV of MRA and CTA for detecting significant stenoses were 92%, 95%, 73%, 99% and 96%, 92%, 67%, 99%, respectively, on assessable segments basis.
Conclusion
Combined with dedicated 32-channel coils, parallel imaging with higher acceleration factors allows significant reduction in imaging time when compared to our previous results. These improvements resulted in substantially improved overall accuracy of MRA in detecting CAD if only assessable segments were considered. Both MRA and CTA have higher negative predictive values and would be the preferred techniques for noninvasive coronary imaging in ruling out CAD.
References
Qi Yang, Kuncheng Li, Li D et al: JACC. 2009, 54: 69-76.
Pouleur AC, Gerber BL et al: Circ Cardiovasc Imaging. 2008, 1: 114-121. 10.1161/CIRCIMAGING.107.756304.
Niendorf T et al: Magn Reson Med. 2006, 56: 167-176. 10.1002/mrm.20923.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Li, K., Yang, Q., Ma, H. et al. Contrast enhanced coronary MRA at 3.0 T using a 32 channel coil to detect coronary artery stenosis - how does it measure up against 64-slice coronary CTA and X-ray angiography?. J Cardiovasc Magn Reson 12 (Suppl 1), P52 (2010). https://doi.org/10.1186/1532-429X-12-S1-P52
Published:
DOI: https://doi.org/10.1186/1532-429X-12-S1-P52
Keywords
- Suspected Coronary Artery Disease
- Significant Stenos
- High Acceleration Factor
- Grappa Factor
- Assessable Segment