Table 1 Differences between CMR measurements and actual results with correlation coefficients in explanted canine hearts

From: Comparison of long and short axis quantification of left ventricular volume parameters by cardiovascular magnetic resonance, with ex-vivo validation

 

Reference (mold)

LAX

SAX 10 mm

SAX 8 mm

SAX 5 mm

LV Volume

23.3 ± 5.3 ml

-2.5 ± 1.1%

CV = 1.18

-7.0 ± 7.8%

CV = 3.63

-0.7 ± 4.6%

CV = -12.1

+0.3 ± 5.3%

CV = -12.7

LA Volume

11.8 ± 7.5 ml

-13.3 ± 7.0%

CV = 1.61

-4.2 ± 6.9%

CV = 3.49

-3.3 ± 6.9%

CV = 6.07

+5.3 ± 9.1%

CV = -9.24

LV Mass

99.6 ± 5.5 g

-3.3 ± 3.0%

CV = 2.68

+7.5 ± 5.6%

CV = -3.49

+4.0 ± 4.2%

CV = -5.31

+0.1 ± 6.0%

CV = 17.3

  1. Difference CV shows the degree of variation by providing the ratio of the standard deviation relative to the mean. Values are expressed as mean ± SE. LV: Left ventricular; LA: Left atrial; LAX: Long axis method; SAX: Short axis method; CV: coefficient of variation of the difference between mold and CMR values.