Table 1 MRI in Patients with Pacemakers and ICDs.

From: Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR Conditional" designs

Author

Device

Year

Patient/Studies Report Type

MRI

Condition

Findings

Iberer [8]

PPM

1987

1/1 Case

 

No adverse effect

Alonga [11]

PPM

1989

1/1 Case Intentional

1.5 T

Brain

No adverse effect

Inbar [12]

PPM

1993

1/1 Case Intentional

1.5 T

Brain

No adverse effect

Gimbel [15]

PPM

1996

5/5 Retrospective Intentional

0.35-1.5T

Cardiac, Brain, C-Spine

Two second pause

Garcia-Boloa [16]

PPM

1998

1/2 Case Intentional

1.0 T

Brain

No adverse effect

Fontaine [98]

PPM

1998

1/1 Case Intentional

1.5 T

Brain, C-Spine

Rapid pacing

Sommer [143]

PPM

1998

18/18 Prospective

0.5 T

Brain, Cardiac, Vascular

Asynchronous mode due to activation of the reed switch in all patients

Sommer [18]

PPM

2000

45/51 Prospective

0.5 T

Multiple

No adverse effect

Valhaus [20]

PPM

2001

32/34 Prospective

0.5T

Multiple

Decrease in battery voltage recovered at 3 months

Martin [29]

PPM

2004

54/62 Prospective

1.5 T

Multiple

Significant change in pacing threshold in 9.4% of leads, and 1.9% of leads requiring an increase in programmed output.

Del Ojo [32]

PPM

2005

13/13 Prospective

2.0 T

Multiple

No adverse effect.

Rozner [41]

PPM

2005

2/2 Case Intentional

1.5 T

Thorax, Lumbar

Transient change to ERI in 1 patient.

Gimbel [33]

PPM

2005

10/11 Prospective

1.5 T

Brain

C-Spine

Small variances in pacing threshold were seen in four patients.

Sommer [50]

PPM

2006

115/82 Prospective

1.5 T

Extra-thoracic

Significant increase in pacing threshold, decreased lead impedance, and decrease in battery voltage. No inhibition of pacing or arrhythmias and no leads which required an increase in pacing output.

Heatlie [52]

PPM

2007

5/6 Prospective

0.5 T Cardiac

Pacing at maximum voltage at a fixed rate of 100 beats/minute in one patient.

Anfinsen [22]

ICD

2002

1/1 Case Inadvertent

0.5 T

Brain

Inappropriate sensing, battery voltage transient change to EOL.

Fiek [28]

ICD

2004

1/1 Case Inadvert

0.5 T

Brain

Unable to communicate with device.

Coman [30]

ICD

2004

11/11 Prospective

1.5 T

Cardiac, Vascular, General

Brief asymptomatic pause in 1 patient.

Unable to communicate with device in 1 patient.

Gimbel [34]

ICD

2005

7/8 Prospective

1.5 T

Brain, L-Spine

"Power on reset" electrical reset requiring reprogramming in 1 patient.

Roguin [39]

ICD

2005

1/1 Case Intentional

1.5 T Cardiac

No adverse effect.

Wollmann [44]

ICD

2005

1/3 Case Intentional

1.5 T

Brain

No adverse effect.

Naehle [46]

ICD

2006

1/1 Case Intentional

1.5 T

Brain

No adverse effect.

Nazarian [47]

PPM 31

ICD 24

2006

68/55 Prospective

1.5 T

No adverse effect.

Nemec [48]

ICD

2006

1/1 Case Unintentional

Not specified

Brain

Noise detected as ventricular tachycardia and ventricular fibrillation, with no therapy presumably due to magnetic mode activation. Asynchronous pacing due to noise-reversal mode.

Sardanelli [144]

PPM

2006

1/1 Case Intentional

1.5 T

Breast

No adverse effect.

Mollerus [56]

PPM 32

ICD 5

2008

37/40 Prospective

1.5 T

Truncal, non-truncal

No adverse effect.

No changes in cardiac troponin-I

Naehle [57]

PPM

2008

44/51 Prospective

3.0 T

Brain

No adverse effect.

No changes in cardiac troponin-I

(Use of transmit-receive head coil)

Gimbel [59]

PPM

2009

1/1 Case Intentional

2.0 T

Brain

Asystole

Goldsher [60]

PPM

2009

1/1 Case Intentional

1.5 T

Cervical

No adverse effect.

Scan one day after implant

Pacemaker dependent

Mollerus [61]

PPM 46

ICD 6

2009

52/59 Prospective

1.5 Truncal, non-truncal

MRI-related ectopy 7 pt

Naehle [62]

PPM

2009

47/171 Case Intentional

1.5 T

General

Statistically significant but clinically irrelevant change in pacing capture threshold and battery voltage.

2 or more serial scans.

Pulver [64]

PPM

2009

8/11 Prospective

1.5 T

Cardiac, non-cardiac

No adverse effect.

Congenital heart disease with 9 epicardial leads.

Strach [67]

PPM

2010

114/114 Prospective

0.2 T

General

No adverse effect

Millar [66]

PPM

2010

1/1

1.5 T

Brain C-spine

No adverse effects

Burke [117]

PPM 24

ICD 10

CRT ICD 4

2010

38/92 Prospective

1.5 T

Brain, Spine, Pelvis, Extremity

No adverse effects

No changes defibrillation threshold (ICD)

Buendia [100]

PPM 28

ICD 5

2010

33/33 Prospective

1.5 T

Cardiac Brain, Spine,

Abdominal, Extremity

Temporary communication failure in two patients.

Sensing errors during imaging in two patients.,

Safety signal generated in one pacemaker at the maximum magnetic resonance frequency and output level.

Cohen [70]

PPM 74

ICD 31

2010

105/105 Prospective

1.5 T

No deaths, device failures, genrerator/lead replacements, loss of capture, or electrical reset.

Decrease in battery voltage of ≥ 0.04 V in 2%

Lead impedance change ≥ 50 Ohms in 3%

High voltage impedance change ≥ 3 Ohms in 10%

Decrease in R wave amplitude in 2%

Pacing threshold increase of ≥ 0.5 V at 0.4 ms in 1% of leads

Wilkoff [81]

MR ConditionalPPM

2011

226/226 Prospective

1.5 T

Brain and lumbar

No adverse effect

Quarta [140]

MR ConditionalPPM

2011

1/1 Prospective

1.5 T

Brain, Cardiac

No adverse effect

  1. Adapted and Updated from Shinbane et al. 2007. [1]
  2. Case = case report, EOL = end-of-life, ERI = elective-replacement indices; ICD-implantable cardiac defibrillator; PPM = pacemaker; T = Tesla,.