Skip to main content

Table 1 MRI-results

From: Accuracy of aortic pulse wave velocity assessment with velocity-encoded MRI: validation in patients with Marfan syndrome

 

PWV 2-slice method

PWV 4-slice method

Pearson r

0.47

0.82

Mean difference ± standard deviation (m/s)

0.40 ± 1.40

0.18 ± 0.56

p-value t-test

0.24

0.19

Mean unsigned error

18%

8%

Coefficient of variation

25%

10%

Mean total scan time

8 minutes

2 minutes