Figure 3 | Journal of Cardiovascular Magnetic Resonance

Figure 3

From: Predicting hemodynamics in native and residual coarctation: preliminary results of a Rigid-Wall Computational-Fluid-Dynamics model (RW-CFD) validated against clinically invasive pressure measures at rest and during pharmacological stress

Figure 3

Bland-Altman plots: Comparison of the clinically invasive and the predicted CFD pressure gradient (mmHg) across the aortic coarctation at rest (A) and isoprenaline pharmacological stress (B). The dot-dashed grey horizontal lines represent the mean difference between CFD and invasive data (MB, mean bias), and the paired dotted red horizontal lines represent ±2 standard deviations from this mean difference (LOA), 95% limits of agreement). Note the increased bias during the stress condition compared to the good agreement at rest condition.