Detection of myocardial oedema with the use of diffusion-weighted imaging in acute myocardial infarction
Journal of Cardiovascular Magnetic Resonance volume 13, Article number: P98 (2011)
Introduction
MR diffusion-weighted imaging is an important application for oedema detection in various tissues. Evaluation of the area at risk in reperfused acute myocardial infarction (AMI) is currently performed with STIR T2-weighted and LGE sequences.
Purpose
The aim of the study was to find practical application for a new diffusion-weighted sequences in evaluation myocardial oedema and compare it with routinely used STIR-T2 techniques.
Methods
In preliminary study myocardial oedema in 15 patients (13 male) with STEMI within 2-4 days were evaluated .The CMR examination was performed on a 1,5 T scanner (Magnetom Avanto; Siemens; Germany; Erlangen;) using a 8-channel phased-array coil. The parameters of the diffusion-weighted EPI sequence (DWI) were as follows: slice thickness 10mm, repetition time (depending on patient breath cycle) 3-4s, echo time 78ms, bandwidth 1,736 Hz/Px. The DW sequence was ECG-gated and synchronized to the respiratory cycle using PACE technique. Each slice was acquired with b = 50 s/mm2, 400s/mm2 and 800 s/mm2 with three perpendicular directions of the diffusion gradient. DW, STIR T2-weighted and LGE images were obtained in 2-chamber, 4-chamber or short-axis planes. Images were analysed quantitatively, contrast to noise ratio (CNR) of high signal (oedema) to healthy myocardium (CNR1) and high signal to blood (CNR2) were calculated. For statistical analysis a non parametric Wilcoxon test with significance level of p<0,05 was used.
Results
The CNR were significantly higher on DWI than on STIR T2-weighted images: CNR1 (22±7 vs 12±8 p= 0,004, respectively)and CNR2 (28±10 vs 21±9, p=0,02 , respectively).
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Kociemba, A., Lanocha, M., Katulska, K. et al. Detection of myocardial oedema with the use of diffusion-weighted imaging in acute myocardial infarction. J Cardiovasc Magn Reson 13 (Suppl 1), P98 (2011). https://doi.org/10.1186/1532-429X-13-S1-P98
Published:
DOI: https://doi.org/10.1186/1532-429X-13-S1-P98
Keywords
- Acute Myocardial Infarction
- Myocardial Oedema
- Slice Thickness 10mm
- Breath Cycle
- Reperfused Acute Myocardial Infarction