Figure 11 | Journal of Cardiovascular Magnetic Resonance

Figure 11

From: Cardiovascular magnetic resonance physics for clinicians: part II

Figure 11

Choice of imaging pulse sequence for myocardial perfusion MRI. Diagram illustrating the choices of data acquisition pulse sequence for myocardial perfusion MRI. A saturation recovery preparation pulse is employed to ensure T1-weighting followed after a delay by the image data acquisition pulse sequence. The most common data acquisition pulse sequences used are fast/turbo spoiled gradient echo (FGE), balanced steady state free precession (bSSFP) or hybrid (or segmented) EPI. The number of slices that can be acquired (in this case three) is limited by the R-R interval, the saturation delay and the length of data acquisition period.