Figure 3 | Journal of Cardiovascular Magnetic Resonance

Figure 3

From: Reproducibility of in-vivo diffusion tensor cardiovascular magnetic resonance in hypertrophic cardiomyopathy

Figure 3

cDTI data set for initial and repeat scan in the least reproducible example, most affected by artefacts data set of the ten investigated in this study. The three cDTI slices are depicted as b0 images after averaging, with the derived colourised FA, MD and HA maps. The bottom images show the matching LGE slices with arrows highlighting the enhanced regions. The plots show the measured FA and MD values for the myocardial segment regions only (the blue lines represent the healthy mean value and the ± 2SD limits reported by Nielles-Vallespin et al.,[22] and the error bars in the data points represent the standard deviation measured in each LV wall). Differences can be seen in FA, MD and HA maps, particularly in the mean diffusivity of the septal region mid-slice due to a signal loss artefact in the b0 images of the repeat scan. The HA distribution maps shows noisy pixels due to motion artefacts which are not reproducible. No definite pattern of change is seen in the area of LGE in the anterior wall or elsewhere.