Does change in heart rate and blood pressure during adenosine stress perfusion cardiovascular magnetic resonance (A-CMRP) imaging predict perfusion defects?
Journal of Cardiovascular Magnetic Resonance volume 14, Article number: M5 (2012)
Summary
Peripheral haemodynamic response during adenosine stress CMR perfusion does not vary among those with or without perfusion defects identified during routine clinical practice.
Background
Haemodynamic response to adenosine infusion during CMR stress perfusion continues to be used as surrogate to determine adequacy of hyperaemia in clinical setting. Our aim was to determine the proportion of patients who achieve pre-specified adequate response to adenosine (increase in heart rate (HR) by ≥ 10 beats or a drop in systolic blood pressure (SBP) by 10mm Hg from baseline reading) and if this varies in patients identified with and without perfusion defect in a clinical setting.
Methods
We prospectively identified 94 consecutive patients undergoing A-CMRP, 7 of who were excluded because of missing data. 87 patients underwent first pass CMRP on a Philips Achieva CV 1.5 T MR scanner (Philips, The Netherlands), with standardised SCMR acquisition protocol infusing adenosine at 140 μg/kg/min for 3 minutes with HR and SBP recorded at baseline, one, two and three minutes. All patients confirmed abstinence from caffeine and chocolate for >24 hours and no patients were on theophylline or dipyridamole. Descriptive statistics and logistic regression were performed.
Results
67 (77%) patients achieved the pre-specified adequate haemodynamic response and this was mainly driven by increase in heart rate.
The OR (adjusted for age) for predicting perfusion defect by adequate haemodynamic response in patients with and without CAD was 2.5 (0.58, 10.40, p=0.21) and 3.26 (0.48, 22.13, p 0.20) respectively.
Conclusions
Our study shows that in patients undergoing adenosine (140 μg/kg/min) stress perfusion, there is no significant difference in peripheral haemodynamics in patients identified with or without perfusion defect and challenges their use as a marker of hyperaemic response. It remains to be seen whether high dose adenosine will influence these findings and change clinical practice.
Funding
No funding was obtained for this study.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Rathod, K.S., Davies, C., Petersen, S.E. et al. Does change in heart rate and blood pressure during adenosine stress perfusion cardiovascular magnetic resonance (A-CMRP) imaging predict perfusion defects?. J Cardiovasc Magn Reson 14 (Suppl 1), M5 (2012). https://doi.org/10.1186/1532-429X-14-S1-M5
Published:
DOI: https://doi.org/10.1186/1532-429X-14-S1-M5