Table 5 T1-weighted saturation-recovery gradient echo for first pass myocardial perfusion

From: Guidelines and protocols for cardiovascular magnetic resonance in children and adults with congenital heart disease: SCMR expert consensus group on congenital heart disease

  Infant/small child Large child/adult
In-plane resolution (mm) 1.5-2.0 2.0-2.5
Slice thickness (mm) 5-8 8-10
Image acquisition timing 1 R-R or 2 R-R 1 R-R