Table 1 CNR values between the blood in the ventricles and the myocardial wall

From: Cardiac-respiratory self-gated cine ultra-short echo time (UTE) cardiovascular magnetic resonance for assessment of functional cardiac parameters at high magnetic fields

CNR§

End-diastole$

End-systole$

UTE (left ventricle – left myocardium)

22.7 ± 5.3

21.1 ± 4.5

UTE (right ventricle – right myocardium)

25.5 ± 7.8

21.5 ± 5.2

FLASH (left ventricle – left myocardium)

40.8 ± 12.0

36.7 ± 11.0

FLASH (right ventricle – right myocardium)

39.6 ± 11.0

33.1 ± 9.7

  1. §Contrast-to-noise ratio (CNR) was measured in the end-diastolic and end-systolic phase by self-gated FLASH and self-gated UTE sequences.
  2. $Values are given as mean ± standard deviation values, averaged over six healthy C57BL/6 mice.