Table 3 IR-SE, MOLLI T1 and partition coefficient values

From: Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners

  T1 (ms) Partition coefficient (%)
  Pre-contrast vials Post-contrast vials  
  “Myocardium” “Blood Pool” “Myocardium” “Blood Pool”  
1 869 (773) 1408 (1333) 450 (416) 316 (322) 43.7 (47.2)
2 892 (814) 1434 (1369) 468 (435) 331 (337) 43.7 (47.9)
3 1528 (1332) 1894 (1750) 603 (575) 367 (391) 45.8 (49.2)
4 1643 (1253) 2041 (1693) 649 (593) 398 (423) 46.0 (49.8)
5 1537 (1272) 1882 (1630) 606 (583) 370 (400) 46.2 (50.0)
  T2 (ms)  
1.5T 40 290 53 179  
3T 51 273 49 166  
  1. Note: IR-SE T1 and partition coefficient values are shown in bold. MOLLI T1 and partition coefficient values are displayed inside parenthesis. Data obtained from 4 vials that mimic the T1/T2 of pre- and post-contrast myocardium and blood pool from the 5 scanners. T2 values for both 1.5T and 3T are displayed at the bottom of the table.