Regional variations in T1 in the healthy left ventricle
Journal of Cardiovascular Magnetic Resonance volume 15, Article number: E43 (2013)
Background
T1 mapping is a newer technique for characterizing the myocardium, however regional variations in the healthy left ventricle are not yet fully explored, complicating interpretation of observed pathologic differences.
Methods
We assessed regional variation in T1 in the septum of patients with no left ventricular pathology. An adjunct phantom study was performed to assess whether location closer to a surface coil affected T1. Main study: Retrospective. 40 patients whose cardiac MRI showed no left ventricular disease were identified. Patients were imaged on a Siemens Avanto 1.5T scanner with a 32-channel cardiac coil. All patients had a 4-chamber TI scout approximately 12 minutes post injection of 0.14 mmol/kg gad-BOPTA. QMASS (Medis, NL) was used to generate T1 maps and measurements obtained in small regions of interest throughout the septum. Basal, middle third, and apical regions were evaluated in all 40; a subset of 15 patients had an additional roi placed halfway between the middle and apical ROIs. Adjunct study: TI scout and MOLLI images of a liquid phantom (homogeneous T1 of approximately 315 ms, NiSO4 in H2O, NaCl) phantom were obtained at several simulated heart rates.
Results
Measured T1 values in the septum were: Basal septum 384 ±46 ms (287-484). Midventricular septum 388 ±44 ms (271-477). ¾ distance base to apex along the septum, 356 ±56 ms (233-456). Apical septum 346 ±52 ms (197-459).
By 2-tailed paired t-test, T1 of the basal vs. mid-ventricular septum was not significantly different (p=0.33). However, mid-ventricle was different from ¾ (p=0.008) and apical septum (p<0.0001) with paired t-tests. Unpaired t-test comparison of midventricular to apical septum showed p=0.0002 likelihood that the difference was due to chance.
In the homogeneous T1 liquid phantom, the intra-study T1 difference (i.e., heterogeneity by region) derived from TI scout imaging was 7-15 ms, from MOLLI 10-35 ms. There was thus less intra-study regional variation using the TI scout. However, whole-phantom T1 measured by TI scout was more affected by heart rate (288-314 ms) than MOLLI (314-316 ms).
Conclusions
1) We found a gradient of T1 over the distal half of the septum. The findings may reflect a systematic artifact, or a true regional difference in septal myocardial T1. In either case, the implication is that exact slice location may be important when performing T1 mapping.
Work is currently underway to explore whether this finding reflects a true difference in regional myocardium, or an artifact. Possible explanations include: regional differences in extracellular volume, collagen, or fat content of the normal septum, effects of fiber orientation, regional perfusion / gadolinium kinetics, volume averaging of blood pool related to curvature of the distal septum, motion effects, and susceptibility artifacts related to proximity to lung vs mediastinal tissues.
2) Phantom studies excluded the possibility that this variation is due simply to the fact that the apex is closer to the receiver coil.
Funding
none
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Rovno, H.D., Akers, S. & Litt, H. Regional variations in T1 in the healthy left ventricle. J Cardiovasc Magn Reson 15 (Suppl 1), E43 (2013). https://doi.org/10.1186/1532-429X-15-S1-E43
Published:
DOI: https://doi.org/10.1186/1532-429X-15-S1-E43