Analysis of cyclic deformation of the aortic annulus by cine-MRI
Journal of Cardiovascular Magnetic Resonance volume 15, Article number: P277 (2013)
Background
Imaging of the aortic annulus is paramount for the success of transcatheter aortic valve implantation (TAVI) to obviate prosthesis size mismatch. The aortic annulus is a dynamic structure with alterations of its ellipsoid shape during the cardiac cycle. The objective of this study was to investigate with cine-MRI the changes of the sagittal and coronal diameter of the aortic annulus and whether the point in time of measurement is relevant for prosthesis size selection.
Methods
40 patients (17 males, 84.3 ± 6.7 years) with severe aortic stenosis were studied with a 1.5 Tesla MR Scanner (Achieva, Philips, The Netherlands) before TAVI. Cine-MRI were acquired in coronal axis and sagittal (3-chamber) view of the left ventricle and the aortic annulus diameter was measured at maximum systole and minimal diastole in each view and the mean annulus diameter was calculated for each phase. Additionally an eccentricity index (EI), the ratio of maximum and minimum diameter was calculated as a measure of circularity of the annulus.
Results
The maximum diameter of the aortic annulus was measured at systole for both sagittal (mean 23.7±1.7 mm) and coronal axis (mean 25.0±1.9 mm), as the minimal diameter was measured at diastole for the sagittal (mean 21.1±1.9 mm) and the coronal view (mean 23.6±2.4 mm). The degree of deformation was apparently greater in the sagittal direction than in the coronal extent (2.6±0.6 mm vs. 1.5±0.8 mm, respectively, p<0.01). This resulted in a more circular shape of the annulus during systole, reflected by a decreased EI in comparison to diastole (EI=1.06±04 vs. EI=1.12±0.08, respectively, p<0.01). Also the mean annulus diameter was calculated 2 mm smaller than in systole (22.33 ±2.05 mm vs. 24.34±1.75 mm).
Conclusions
The deformation of the annulus during the cardiac cycle, specifically in the sagittal dimension results in diameter variations of at least 2 mm. It is reasonable that the assessment of the annulus size should be performed in systole, when the annulus is also more circular than in diastole.
Funding
No disclosures.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Boering, Y.C., Gruenig, S., Bönner, F. et al. Analysis of cyclic deformation of the aortic annulus by cine-MRI. J Cardiovasc Magn Reson 15 (Suppl 1), P277 (2013). https://doi.org/10.1186/1532-429X-15-S1-P277
Published:
DOI: https://doi.org/10.1186/1532-429X-15-S1-P277