Skip to main content

Table 3 Pacemaker statistics

From: Safe performance of magnetic resonance of the heart in patients with magnetic resonance conditional pacemaker systems: the safety issue of the ESTIMATE study

Parameter

Pre MR FU (n = 36) [median]

3-month FU (n = 36) [median]

15-month FU (n = 36) [median]

p = *

Time from MR (months)

-

2.8 ± 0.7 [3.0]

16.0 ± 3.2 [15.2]

-

Atrial pacing (%)

40 ± 37 [23]

42 ± 36 [25]

45 ± 36 [40]

n.s.

Ventricular pacing (%)

26 ± 39 [2]

26 ± 38 [3]

30 ± 38 [6]

n.s.

AF-Burden

17 ± 34 [0]

15 ± 33 [0]

20 ± 37 [0]

n.s.

  1. FU…follow-up.
  2. *Oneway ANOVA.