Background
Adenosine perfusion cardiac magnetic resonance imaging (CMR) at 3 Tesla has been suggested to yield superior diagnostic accuracy in comparison to 1.5 Tesla. We sought to evaluate its diagnostic accuracy in a patient cohort with intermediate risk and typical angina symptoms in comparison to quantitative coronary analysis (QCA) and fractional flow reserve (FFR).
Methods
Eighty-six patients with intermediate risk, angina pectoris, and suspected coronary artery disease were enrolled into the study. All patients were examined in a 3 T CMR system including functional imaging of the left ventricle, adenosine and rest perfusion imaging using a bolus of 0.075 mmol/kg gadolinium based contrast agent (Dotarem, Guerbet, France), respectively. CMR studies were performed within 1 week prior to invasive coronary angiography which included QCA and FFR. Myocardial perfusion reserve (MPR) was calculated and correlated to FFR. Visual analysis of the perfusion sequences were compared to QCA. All measurements were performed by two blinded and experienced observers in consensus.
Results
Visual CMR analysis resulted in a specificity of 0.88 and sensitivity of 0.96 for the detection of a coronary artery stenosis ≥70% as quantified by QCA on a per patient basis. ROC analysis for MPR yielded specificity of 1.0 and sensitivity of 0.9 for the detection of a coronary artery with reduced FFR (≤0.8) on a per vessel basis.
Conclusions
Qualitative and quantitative analysis of adenosine perfusion CMR at 3 Tesla are able to detect significant coronary artery disease with high diagnostic accuracies, respectively. 3 Tesla CMR might thus serve as an important non-invasive diagnostic tool in clinical work-up of the evaluation of coronary artery disease.
Funding
Partly funded by a research grant from Guerbet.
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Buckert, D., Rasche, V., Rottbauer, W. et al. Quantitative and qualitative adenosine perfusion magnetic resonance imaging for the detection of coronary artery disease at 3 Tesla. J Cardiovasc Magn Reson 16 (Suppl 1), P180 (2014). https://doi.org/10.1186/1532-429X-16-S1-P180
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-P180
Keywords
- Cardiac Magnetic Resonance
- Fractional Flow Reserve
- Invasive Coronary Angiography
- Perfusion Magnetic Resonance Imaging
- Myocardial Perfusion Reserve