Table 1 showing the variability during a 52 week period of the Phantom T1 and T2 values and ECV using an 11-cycle 8-image prototype MOLLI. There were 2 phantoms for each T1 and T2 value prepared in standardised laboratory and the value used here represents the average of the two. The phantom ECV was calculated using a hematocrit of 0.43.

From: Longitudinal stability of gel T1 MRI Phantoms for quality assurance of T1 mapping

 

T1

T2

 

Mean (ms)

SD

CoV (%)

Mean (ms)

SD

CoV (%)

Mean (ms)

SD

CoV (%)

 

High Resolution (75bpm)

Low resolution (100bpm)

Spin-echo protocol

Native Myocardium

972

12.6

1.3

972

13.6

1.4

56

1.3

2.2

Native Blood

1579

56.4

4.6

1550

61.9

4.0

234

8.0

3.4

Post Gd Myocardium

498

5.1

1.0

497

13.5

1.4

48

0.9

1.9

Post Gd Blood

372

6.4

1.7

372

6.5

1.8

149

3.0

2.0

ECV

27.4%

0.9

3.2

27.6%

0.9

3.1

  Â